Our Category

Distillation Apparatus Glass Quartz Single Stage Automatic Water-Cabinet

Home    >>    Distillation Apparatus Glass Quartz Single Stage Automatic Water-Cabinet

Product Code : AC-466-45/46

Distillation Apparatus Glass Quartz Single Stage Automatic Water-Cabinet


Price of Product

MRP : 0 /- Rs

Price : /- Rs.

Description

Compact & User Friendly

This is a fully automatic water distillation equipment and can be used in research laboratories, hospitals etc. It is manufactured from QUATRZ. The distillation has following features :

  • Replaceable Silica Sheathed (Quartz) Heater : The horizontal jacket is fitted with removable silica sheathed heater and screw threaded connectors for ease of cleaning.
  • Distillate temperature is approx. between 25o to 45o C which is ensured by an efficient condenser.
  • The stand is powder coated for rust free operation and lustrous appearance.
  • Cooling water from the condenser is re-circulated into the boiling chamber, hence saving energy.
  • Screw Threaded Connectors : These are provided for easy plumbing & Installation to avoid any glassware breakage during assembly.
  • Fuse is provided for safety in case of voltage fluctuation or short circuit.
  • Boiler is provided with Replaceable Constant Level Device for easy draining and cleaning of boiling chamber.
  • The front cover is easily removable making the access to the glass parts very easy.

QUARTZ Glass Single Stage

Model No.

AC-466-45

AC-466-46

Output Cap. Approx

2 ltrs/hr

4 ltrs/hr

Stage

Single Stage

Single Stage

Distillate Quality

Conductivity uS/cm

0.5-0.8

0.5-0.8

Cooling Water Requirement app.

1.5 ltr/min

2.3ltr/min

Power Rating-Total

1.7 KW

3.0 KW

Reservoir Level Sensor :

This helps in automatic working of the water distillation equipment. Depending on the water level in the reservoir it will automatically switch On/Off the equipment, thus saving energy. (Reservoir can also be provided with a storage capacity of 5 ltr, 10 ltr& 20 ltr made of Borosilicate Glass or polypropylene).at extra cost

Additional Built-in Safety Features

Gate Valve : Conserves cooling water wastage in case of power failure.
Flow Switch : In-built safety feature for unit shut off in case of cooling water failure.
Temperature Switch : The boiler is equipped with a safety temperature switch which shuts off the unit if water level in the boiler falls below the heater, thus saving the glass parts from any damages.

Level Sensor : It shuts off the distillation equipment if the water level in the boiler falls below the heater.